염수 분무 환경 챔버의 일반적인 결함
The enterprise users are often at a loss when dealing with common faults in the salt spray environmental chamber. For this reason, Wuxi LNEYA, a manufacturer of the salt spray environmental chamber, specially reorganized the related common fault treatment technologies of the salt spray environmental chamber.
Salt spray environmental chamber for sale
The salt spray environmental chamber cannot be heated
1. Heat the solid state relay, if there is no output, replace the solid state relay.
2. If the external dial has no output, replace the external dial.
3. If there is heating tube damage, check the heating tube resistance, replace the resistance if there is something abnormal.
4. Lack of water can cause heating shielding, check whether the water level is normal or not and whether the floating ball malfunctions or not.
5. If there is a malfunction of the over-temperature maintenance system, adjust the maintenance temperature value. If the fault still cannot be solved, replace the over-temperature maintenance system.
The heating temperature is abnormal
1. If the set temperature of the temperature controller of the environmental chambers for sale is too low, set the temperature controller at the required temperature.
2. If the set temperature of the safety maintenance switch of the environmental chamber is too low, set the safety maintenance switch at the required temperature.
3. If the set temperature of the temperature controller of the full bucket is too low, set the temperature controller at the required temperature.
4. If the set temperature of the safety maintenance switch of the full bucket is too low, set the safety maintenance switch at the required temperature.
The external dial doesn’t display
1. Check whether or not there is a power supply in the salt spray environmental chamber. If not, ask a professional electrician for a check.
2. Check whether or not the power switch is damaged. If damaged, replace the power switch.
3. Check whether or not the power circuit breaker is damaged. If damaged, replace the circuit breaker.
4. Check whether or not the external dial terminal has a power supply. If there is a power supply, it indicates that the external dial is damage, replace the external dial.
5. Check whether or not the power supply wiring terminal of the salt spray environmental chamber is blackened. If it is blackened, it indicates the possibility of burning out, replace the external dial.
If the failures the enterprise users are encountered are more serious than above, it is suggested to contact the salt mist environmental chamber manufacturer to solve the failure.
관련 권장 사항
-
RF chip high and low temperature automatic test machine control system description
1367Based on the continuous development of the semiconductor chip industry, LNEYA RF chip high and low temperature automatic test machine is also promoted to consumers. So, what do you know about the RF chip high and low temperature automatic tester s...
세부 정보 보기 -
고온 및 저온 습열 테스트 장치의 압축기에 대한 설명
1318고온 및 저온 습열 테스트 장치용 컴프레서의 종류에 따라 가격에 영향을 미치는 경우가 많습니다. 선택할 때 가능한 한 저에너지 냉동 압축기를 선택해야합니다. 이제 중국은 점점 더 많은 관심을 기울이고 있습니다 ...
세부 정보 보기 -
All analysis of refrigerant changing process of industrial chiller unit
1051Refrigerant is an indispensable part of the refrigeration system of the industrial chiller. During the operation of the industrial refrigeration unit, the refrigerant constantly changes in the industrial refrigeration unit to achieve the cooling e...
세부 정보 보기 -
유리 반응기에 대한 냉각 및 가열 시스템의 영향
1248유리 반응기는 부피가 작고 구조가 명확하며 유연한 방식으로 직렬 및 병렬로 편리하게 작동하여 다양한 공정 상황을 완벽하게 충족시킬 수 있으며 반응 경로에 금속 접촉이 없으며 인라인 ...
세부 정보 보기