Failures of Heating Cooling Circulators
The heating cooling circulator can be used in many chemical pharmaceuticals and laboratories. Then, how to solve the failures of the heating cooling circulator?
Failure of circulating pump
The failure of the circulating pump is generally the heat overload of the circulating pump. There are 4 main reasons for this failure:
1. The external circulation pipeline is too long. It is suggested to shorten the external circulation pipeline, reduce the bending of the external circulation pipeline, and then click the Reset button of the electric cabinet surface heat relay;
2. The circulating pipe is blocked. It is suggested to check the degree of cleanliness of the circulating pipe, install a filter, and then click the Reset button of the electric cabinet heat relay;
3. The circulation pump voltage is too low. It is suggested to measure whether the circulation pump voltage is in line with the prescribed range, improve the voltage environment, and then click the Reset button of the electric cabinet heat relay;
4. The regulating current of the heat relay is too small. It is suggested to measure the pump current, compare the heat relay range, turn up the heat relay current properly with the cross screwdriver, and then click the Reset button of the electric cabinet heat relay.
Failure of sensor
The great possibility of sensor failure is the wiring error. At this time, check the wiring in time, and we should pay attention to two kinds of situations, one is the sensor is short-circuited or the temperature is < -160 ℃, the other is the sensor is open circuited or the temperature is > 349℃.
If the sensor is short-circuited or the temperature is <-160 ℃, it may be the fault of the sensor itself. At this time, change the sensor (there is a standby sensor in the equipment line slot), if the sensor connection is wrong, it is suggested to check the connection condition (the control circuit diagram) to correct.
If the sensor is open circuited or the temperature is > 349 ℃, it should be the fault of the sensor itself. At this time, replace the sensor in time (there is a standby sensor in the equipment line slot). It is necessary to pay attention to whether the temperature is too high and beyond the range. If the equipment is out of electricity, check the power supply of the equipment.
The above two failures of the heating cooling circulator are relatively simple. I believe that you can quickly solve these failures with these tips.
관련 권장 사항
-
Selection of heat transfer oil for the refrigeration and heating system of the reactor
1159In the chemical industry, the reactor is a commonly used container, and temperature control is the main quantity to be controlled. The heating and refrigeration cycle equipment of the reactor is an important factor to ensure product quality. The r...
세부 정보 보기 -
초저온 냉각기 기능
1148초저온 냉각기는 마이크로 컴퓨터에 의해 제어되며 냉동 압축기 및 시스템 구성 요소는 에너지 절약 제품으로 전체 기계의 신뢰성을 크게 향상시킵니다. 이 장치는 다단계 에너지를 실현할 수 있습니다 ...
세부 정보 보기 -
Cooling and heating thermostat for running electronic component testing
1114If enterprises want the cooling and heating thermostats for electronic component testing to operate for a long time without any faults, they need to achieve the following points to maintain stable equipment performance. 1. Suitable working enviro...
세부 정보 보기 -
What is the role of the reactor refrigeration and heating circulator? How to choose?
1084The refrigeration and heating circulator is a device that can provide precise temperature control for the reactor. It can realize heating and cooling functions at the same time. It is suitable for chemical reaction processes that require p...
세부 정보 보기