The heating cooling circulator can be used in many chemical pharmaceuticals and laboratories. Then, how to solve the failures of the heating cooling circulator?
Failure of circulating pump
The failure of the circulating pump is generally the heat overload of the circulating pump. There are 4 main reasons for this failure:
1. The external circulation pipeline is too long. It is suggested to shorten the external circulation pipeline, reduce the bending of the external circulation pipeline, and then click the Reset button of the electric cabinet surface heat relay;
2. The circulating pipe is blocked. It is suggested to check the degree of cleanliness of the circulating pipe, install a filter, and then click the Reset button of the electric cabinet heat relay;
3. The circulation pump voltage is too low. It is suggested to measure whether the circulation pump voltage is in line with the prescribed range, improve the voltage environment, and then click the Reset button of the electric cabinet heat relay;
4. The regulating current of the heat relay is too small. It is suggested to measure the pump current, compare the heat relay range, turn up the heat relay current properly with the cross screwdriver, and then click the Reset button of the electric cabinet heat relay.
Failure of sensor
The great possibility of sensor failure is the wiring error. At this time, check the wiring in time, and we should pay attention to two kinds of situations, one is the sensor is short-circuited or the temperature is < -160 ℃, the other is the sensor is open circuited or the temperature is > 349℃.
If the sensor is short-circuited or the temperature is <-160 ℃, it may be the fault of the sensor itself. At this time, change the sensor (there is a standby sensor in the equipment line slot), if the sensor connection is wrong, it is suggested to check the connection condition (the control circuit diagram) to correct.
If the sensor is open circuited or the temperature is > 349 ℃, it should be the fault of the sensor itself. At this time, replace the sensor in time (there is a standby sensor in the equipment line slot). It is necessary to pay attention to whether the temperature is too high and beyond the range. If the equipment is out of electricity, check the power supply of the equipment.
The above two failures of the heating cooling circulator are relatively simple. I believe that you can quickly solve these failures with these tips.
Related recommendations
-
Miniature high-pressure reaction kettle refrigeration heating temperature control system
2010The miniature high-pressure reaction kettle refrigeration heating temperature control system is designed with a gas feed unit, a liquid feed unit, a reaction unit and a collection unit. The raw material gas and raw material liquid are mixed and pr...
View details -
How to Solve the Problem that Water 10hp Refrigerator Can’t Produce Water
2008Wuxi guanya has cold plate freezer for sale for the industrial cold treatment.Well,how to solve the problem of water outlet fault of 10hp cold plate freezer?Therefore,I have sorted out the relevant solutions for the water outlet fault of 10hp cold...
View details -
Thermal Oil System for Reactors in chemical/industrial laboratory
2372A thermal oil system for a reactor is a heat transfer system designed to provide precise and controlled heating or cooling to a chemical or industrial reactor. This system uses a heat transfer fluid, typically a high-temperature minera...
View details -
Application and function of explosion-proof constant temperature circulation system
2013Temperature control in drug research and development and chemical experiments, as well as temperature control in small-scale experimental production and industrial production, all require highly dynamic constant temperature controllers. The chemic...
View details
LNEYA Thermo Refrigeration Co.,Ltd