Failures of Heating Cooling Circulators
The heating cooling circulator can be used in many chemical pharmaceuticals and laboratories. Then, how to solve the failures of the heating cooling circulator?
Failure of circulating pump
The failure of the circulating pump is generally the heat overload of the circulating pump. There are 4 main reasons for this failure:
1. The external circulation pipeline is too long. It is suggested to shorten the external circulation pipeline, reduce the bending of the external circulation pipeline, and then click the Reset button of the electric cabinet surface heat relay;
2. The circulating pipe is blocked. It is suggested to check the degree of cleanliness of the circulating pipe, install a filter, and then click the Reset button of the electric cabinet heat relay;
3. The circulation pump voltage is too low. It is suggested to measure whether the circulation pump voltage is in line with the prescribed range, improve the voltage environment, and then click the Reset button of the electric cabinet heat relay;
4. The regulating current of the heat relay is too small. It is suggested to measure the pump current, compare the heat relay range, turn up the heat relay current properly with the cross screwdriver, and then click the Reset button of the electric cabinet heat relay.
Failure of sensor
The great possibility of sensor failure is the wiring error. At this time, check the wiring in time, and we should pay attention to two kinds of situations, one is the sensor is short-circuited or the temperature is < -160 ℃, the other is the sensor is open circuited or the temperature is > 349℃.
If the sensor is short-circuited or the temperature is <-160 ℃, it may be the fault of the sensor itself. At this time, change the sensor (there is a standby sensor in the equipment line slot), if the sensor connection is wrong, it is suggested to check the connection condition (the control circuit diagram) to correct.
If the sensor is open circuited or the temperature is > 349 ℃, it should be the fault of the sensor itself. At this time, replace the sensor in time (there is a standby sensor in the equipment line slot). It is necessary to pay attention to whether the temperature is too high and beyond the range. If the equipment is out of electricity, check the power supply of the equipment.
The above two failures of the heating cooling circulator are relatively simple. I believe that you can quickly solve these failures with these tips.
Related recommendations
-
What are the applications of external circulation oil chillers?
2198External circulation oil chillers are mainly used in the industrial field, especially suitable for situations that require cooling oil to maintain normal equipment operation and extend equipment service life. The following are some typical applica...
View details -
Specific conditions of high and low temperature test of EV (electric vehicle) battery
2614There are mainly 2 types of vehicles:1. Battery for pure electric vehicle2. Fuel cell, dedicated to fuel cell electric vehicles Thermal test: This test is performed to characterize the reaction of cells to the high-temperature environment. a) Adju...
View details -
High-low temperature control system with different configurations of high-pressure reactors
1732Different configurations of high pressure reactors, the quality of products brought by different manufacturers is not the same, but the points of attention are the same, you need to pay attention to it. 1. To be used at the designated location and...
View details -
Analysis of Static Strength of Battery Box in Power Battery Test and Analysis System
2551The battery box of new energy vehicles is one of the core components of electric vehicles. Designing high-strength, high-rigidity, high-security, high-lightweight battery boxes is of great significance to the development of electric vehicles. Over...
View details
LNEYA Thermo Refrigeration Co.,Ltd