Failures of Heating Cooling Circulators
The heating cooling circulator can be used in many chemical pharmaceuticals and laboratories. Then, how to solve the failures of the heating cooling circulator?
Failure of circulating pump
The failure of the circulating pump is generally the heat overload of the circulating pump. There are 4 main reasons for this failure:
1. The external circulation pipeline is too long. It is suggested to shorten the external circulation pipeline, reduce the bending of the external circulation pipeline, and then click the Reset button of the electric cabinet surface heat relay;
2. The circulating pipe is blocked. It is suggested to check the degree of cleanliness of the circulating pipe, install a filter, and then click the Reset button of the electric cabinet heat relay;
3. The circulation pump voltage is too low. It is suggested to measure whether the circulation pump voltage is in line with the prescribed range, improve the voltage environment, and then click the Reset button of the electric cabinet heat relay;
4. The regulating current of the heat relay is too small. It is suggested to measure the pump current, compare the heat relay range, turn up the heat relay current properly with the cross screwdriver, and then click the Reset button of the electric cabinet heat relay.
Failure of sensor
The great possibility of sensor failure is the wiring error. At this time, check the wiring in time, and we should pay attention to two kinds of situations, one is the sensor is short-circuited or the temperature is < -160 ℃, the other is the sensor is open circuited or the temperature is > 349℃.
If the sensor is short-circuited or the temperature is <-160 ℃, it may be the fault of the sensor itself. At this time, change the sensor (there is a standby sensor in the equipment line slot), if the sensor connection is wrong, it is suggested to check the connection condition (the control circuit diagram) to correct.
If the sensor is open circuited or the temperature is > 349 ℃, it should be the fault of the sensor itself. At this time, replace the sensor in time (there is a standby sensor in the equipment line slot). It is necessary to pay attention to whether the temperature is too high and beyond the range. If the equipment is out of electricity, check the power supply of the equipment.
The above two failures of the heating cooling circulator are relatively simple. I believe that you can quickly solve these failures with these tips.
Related recommendations
-
Why do mechanical factories need chillers?
1788There are many reasons why mechanical factories need chillers, and chillers play a very important role in the mechanical processing process, which can improve production efficiency, ensure stable operation of equipment, and extend equipment servic...
View details -
Wuxi Guanya: Focus on R&D of refrigeration heating equipment to help develop new energy
2573On October 17, 2018, the three-day API was held in Nanjing International Expo Center. The exhibition was crowded with people. Wuxi Guanya was a frequent visitor to the API. Wuxi Guanya Constant Temperature Refrigeration Technology Co., Ltd. (herei...
View details -
Way to Check Problems of the Lubricating Oil System of the Screw Chiller Factory
2146Everyone pays more attention to the refrigerant circulation system and the chilled water circulation system of the screw chiller factory operation system. However, lubricating system is often neglected by everyone. In fact, this system of cold wat...
View details -
Refrigeration heating temperature control system equipment used in hydrothermal reactor
1846Refrigeration heating temperature control system can be used in a variety of reactors, reactors, and distillation equipment. Today we will analyze what you need to pay attention to when applying it to a hydrothermal reactor. The heat transfer medi...
View details
LNEYA Thermo Refrigeration Co.,Ltd