How To Choose the Ultra Low Temperature Recirculating Chiller Expansion Valve?
Selection and installation of the expansion valve for ultra low temperature recirculating chillers: the selection of expansion valve plays an important role in the performance of the ultra low temperature recirculating chiller refrigeration system. The correct choice of expansion valve will maximize the use of the evaporator and make the evaporator always be matched to the heat load.
First, the selection of thermal expansion valve.
In the ultra low temperature recirculating chiller refrigeration system, the selection and installation of the expansion valve is very important. Only the correct selection and installation can ensure the normal operation of the unit. Therefore, when selecting the expansion valve, we can use the appropriate size of the thermal expansion valve according to the size of the evaporator load and the evaporating temperature. The capacity of the commonly used expansion valve is about 20% larger than the evaporator heat load. For small chiller refrigeration systems, gas-filled thermal expansion valves are generally used, and medium-sized ultra-low temperature circulating chiller refrigeration systems mostly use adsorption-filled or mixed-filled expansion valves.
Second, the expansion valve installation.
1. The chiller expansion valve is installed in front of the evaporator and is usually installed in a vertical position. Considering the ease of installation and maintenance, the cooling loss is small and should be as close as possible to the inlet of the evaporator.
2. The working condition of the expansion valve for the ultra low temperature recirculating chiller has a direct relationship with the installation position of the temperature sensing package. Because the expansion valve completely adjusts the opening size of the expansion valve by the temperature change of the temperature sensing package. The temperature sensor package is installed on the pipe at the outlet of the chiller evaporator. As the evaporator outlet pipe grows, its superheat gradually increases. Therefore, the position of the expansion valve temperature sensing package is directly related to the amount of fluid supplied from the expansion valve.
Related recommendations
-
60P ultra-low temperature water chiller consumes too much energy and needs attention
189860p ultra-low temperature water chillers of different manufacturers have different performance configurations and different energy consumption. If you find that the energy consumption is too large, you must check in time to see if there is a probl...
View details -
Instructions for use of various components of high and low temperature cycle machine
2216In the high and low temperature cycle machine, each accessory has its own performance advantages. When running, we need to have a clear understanding of each accessory in order to run the device better. The compressor functions to compress and dri...
View details -
Solution to oil spill of reactor cooling and heating circulator
1900When we use some reactor cooling and heating circulators, we will find that the system pressure fluctuates up and down as the temperature rises, and the oil tank is overflowing. So what are the reasons for these conditions in the cooling and heat...
View details -
Explosion-proof ultra-low temperature refrigerator internal leakage instructions
23381. Sound leak detection method: First, the refrigeration pipeline of the explosion-proof ultra-low temperature refrigerator is filled with nitrogen. Then use a stethoscope to check the interior. When you hear a leak in a certain area, mark it. Car...
View details
LNEYA Thermo Refrigeration Co.,Ltd