How to eliminate the high pressure alarm of the industrial ultra-low temperature freezer?
During use, if a fault is not solved in time, it will affect the use of industrial ultra-low temperature freezer. How to solve the problem of high-pressure alarm?
If the water-cooled industrial ultra-low temperature freezer presents a high-pressure alarm, it is mainly related to the cooling water system. The water-cooled industrial ultra-low temperature freezer provides circulating cooling water through a cooling tower. If you forget to open the cooling water valve during use, If the cooling water cannot be circulated to the chiller, it will show high pressure problems. In addition, check whether the cooling water tower is defective and cannot safely and stably deliver the cooling water.
The cooling water flow is too small, or the temperature is too high and the heat dissipation is not good. This problem will also occur. If the cooling water flow is small, it is necessary to check whether the pipe size of the supporting device matches the size of the industrial ultra-low temperature freezer group, or whether the water valve can be opened to a large position; if it is an air-cooled industrial ultra-low temperature freezer, we need to look at the ambient temperature of the device If it can be exceeded, the heat must be exhausted in time to ensure that the industrial ultra-low temperature freezer unit is in an air circulating environment.
Industrial ultra-low temperature freezers are used for too long, and it is easy to accumulate dirt and other debris on the pipe wall or condenser copper pipes, which will inevitably affect the refrigeration cycle system. Therefore, we regularly provide industrial ultra-low temperature freezers according to the maintenance time recommended by the manufacturer Do maintenance cleaning work (cleaning the cooling water system, cleaning the condenser, changing the filter, etc.).
Check whether the refrigerant is excessive. The industrial ultra-low temperature freezer unit will be filled with the appropriate amount of refrigerant before leaving the factory. Therefore, if there is no leakage of the refrigerant, the user does not need to charge the refrigerant. Too much refrigerant will only affect the unit’s failure. use.
Whether the opening of the shrink valve is too small, for this, everyone can gradually adjust the opening of the shrink valve to eliminate the problem.
Related recommendations
-
Motor test bench test and temperature system for vehicle components
2878All vehicle components must undergo rigorous testing, but safety programs must undergo more stringent durability and terminal testing. Many physical methods can be used to load the motor, but the small motor test bench is favored for its high spee...
View details -
Vocs adsorption equipment refrigeration temperature control instructions
2285With the continuous improvement of environmental protection requirements, each factory has also concentrated the processing of vocs gas generated in production. LNEYA vocs adsorption equipment can process various vocs gases. So how much do you kno...
View details -
Discussion on maintenance of high and low temperature integrated cycle machine
2240LNEYA high and low temperature integrated cycle machine is a refrigeration heating equipment produced by the use of refrigeration heating dynamic temperature control technology. The temperature range is from -120 °C to 350 °C. It has superior perf...
View details -
How to improve the efficiency of water cooled chillers?
2480Water cooled chiller is a kind of mechanical equipment that uses water resources for refrigeration. In recent years, water-cooled chiller is more and more widely used at home and abroad, mainly because most modern industrial production needs to us...
View details
LNEYA Thermo Refrigeration Co.,Ltd