How to solve the problem of refrigeration pump circulator pump failure?
Cooling heating devices are required in many chemical pharmaceutical and laboratory applications. Then how to solve the failure of the refrigeration heating device is also concerned about how to solve it?
1. Circulating pump failure
The circulating pump fault is generally the thermal overload of the circulating pump. There are four main reasons for this fault: The first is that the outer circulating pipeline of the heat conducting oil is too long. At this time, it is necessary to shorten the outer circulation pipeline, reduce the bending of the outer circulation pipeline, and then click Equipment electrical cabinet surface thermal relay Reset button; the second is the circulation pipeline blockage, cooling and heating device which is good to tell everyone, check the circulation pipeline cleaning, install the filter, and then click the equipment electrical cabinet surface thermal relay Reset button; The third is that the circulating pump voltage is too low. It is recommended to measure whether the circulating pump voltage meets the specified range, improve the voltage environment, click the Reset button of the equipment electrical cabinet surface thermal relay; the fourth is that the thermal relay regulates the current too small. At this time, it is recommended to measure the pump current. Compare the range of thermal relays, Phillips screwdriver to properly rotate the thermal relay current, click on the equipment electrical cabinet surface thermal relay Reset button.
2. Sensor failure
A great possibility of sensor failure is wiring error. Once the fault is encountered, check the wiring situation. At this time, you should pay attention to two situations, one is sensor short circuit or temperature <-160 °C, and the other is sensor open circuit or temperature > 349 ° C.
If the sensor is short-circuited or the temperature is <-160 °C, it may be the fault of the sensor itself. At this time, it is recommended to replace the sensor (there is a spare sensor in the equipment slot). If the sensor is wired incorrectly, It is recommended to check the wiring (corrected circuit diagram) correction.
If the sensor is open circuit or the temperature is >349°C, it should be the fault of the sensor itself. It is necessary to replace the sensor in time (there is a spare sensor in the equipment slot). At this time, it is necessary to pay attention to whether the temperature is too high and the temperature is too high. Check whether the temperature is exceeded. Range, if the equipment is not powered, pay attention to check the power supply of the equipment.
These two faults of the cooling heating device are still relatively simple. I believe that everyone who has seen the sharing can solve it very quickly.
Related recommendations
-
What is the lack of oil in the 120-degree ultra-low temperature refrigerator?
2255The compressor of the -120 degree ultra-low temperature refrigerator is a complex component that runs at high speed. In order to ensure sufficient lubrication of the moving parts of the compressor crankshaft, bearings, connecting rods, pistons, et...
View details -
LNEYA high and low temperature heating refrigerator maintenance instructions
2279High and low temperature heating refrigerators use circulating water to replace circulating water regularly every month. If heat transfer oil is used, heat transfer oil should be replaced periodically every six months to one year. When the high an...
View details -
The Application of Cold Plate Freezer
1837The refrigeration system of cold plate freezer adopts secondary super-cooler technology with the rapid refrigeration and low limiting temperature. It uses imported brand name semi-hermetic piston compressors, semi-hermetic piston compound compress...
View details -
LNEYA closed cycle refrigerator maintenance instructions
1963LNEYA closed cycle refrigerators should start to pay attention to their maintenance work after a long time of use, then how to carry out maintenance work? How to do maintenance work for it? According to the specific use environment and the frequen...
View details
LNEYA Thermo Refrigeration Co.,Ltd